COULTER LH SLIDEMAKER 6605633

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-06-06 for COULTER LH SLIDEMAKER 6605633 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2787244] Customer reported to beckman coulter, inc. (bec) that the coulter lh 750 slidemaker was giving fluid detector 7 error and multiple other errors. Customer reported that fluid was leaking onto the counter, beneath the instrument. Customer reported that the leaked fluid was clear and partly blue, green tinted. Customer reported that the volume of the leak was less than 20 ml. Customer reported that patient results were not affected. There was no report of any adverse event or injury requiring medical intervention or patient treatment. Bec field service engineer (fse) found holes in the tubings on pinch valves pv5 and pv114 on the sample access reservoir. The fse found the liquid on the counter was clear. The fse found the liquid came from pinch valve pv 5. The fse determine the liquid was diluent and clenz which was used to rinse the sample reservoir lines. The fse re-tubed pv5 and pv114.
Patient Sequence No: 1, Text Type: D, B5


[9831131] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2012-01817
MDR Report Key2600043
Report Source05,06
Date Received2012-06-06
Date of Report2012-05-11
Date of Event2012-05-11
Date Mfgr Received2012-05-11
Device Manufacturer Date2008-03-01
Date Added to Maude2012-10-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOULTER LH SLIDEMAKER
Generic NameSPINNER, SLIDE, AUTOMATED
Product CodeGKJ
Date Received2012-06-06
Model NumberNA
Catalog Number6605633
Lot NumberNA
ID NumberSW VERSION 2D2
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S. KRAEMER BOULEVARD BREA CA 92821623 US 92821 6232


Patients

Patient NumberTreatmentOutcomeDate
10 2012-06-06

© 2026 FDA.report