MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2009-10-02 for GETINGE 9125 UNK manufactured by Getinge Disinfection Ab.
[2789313]
During the installation of a waster disinfector and after applying power to the washer the service rep noticed smoke at the ground connection. The service rep observed that the ground wire insulation had begun to melt.