VANDERLIFT WITH SCALE VANDERLIFT W/SCALE SCALE FG859VS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-05-18 for VANDERLIFT WITH SCALE VANDERLIFT W/SCALE SCALE FG859VS manufactured by Sr Instruments, Inc..

Event Text Entries

[16730537] It was reported to the lift mfr by the nursing home that a malfunction occurred involving a broken hook on the scale attached to the lifter. It was reported that the scale was not in use and no pt or staff were injured. Pictures were sent to the lift manufacturer by the nursing home,
Patient Sequence No: 1, Text Type: D, B5


[17044633] Sr asked for the return of the scale to the lift mfr and they reported that the nursing home was not responding to their request. We then asked the lift mfr if he had any more correspondence with the nursing home and the nursing home's response was no casualties, no one was hurt. The scale "broke". A malfunction was reported. Pictures were sent to the lift mfr by the nursing home, the pictures show the scale with the broken hook laying next to it. Missing from the picture was a screw that is used in the hook and a plastic insert that is added by the lift mfr before shipping to the customer. This plastic insert if not used could have caused the misuse of the scale.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1317256-2012-00003
MDR Report Key2617378
Report Source06
Date Received2012-05-18
Date of Report2012-05-14
Date of Event2012-03-12
Date Mfgr Received2012-04-12
Device Manufacturer Date1999-01-01
Date Added to Maude2012-09-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street600 YOUNG ST.
Manufacturer CityTONAWANDA NY 14150
Manufacturer CountryUS
Manufacturer Postal14150
Manufacturer Phone7166935977
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVANDERLIFT WITH SCALE
Generic NamePATIENT LIFT WITH SCALE
Product CodeFRW
Date Received2012-05-18
Model NumberVANDERLIFT W/SCALE
Catalog NumberSCALE FG859VS
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSR INSTRUMENTS, INC.
Manufacturer Address600 YOUNG ST. TONAWANDA NY 14150 US 14150


Patients

Patient NumberTreatmentOutcomeDate
10 2012-05-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.