FM-15UA-OH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2012-06-12 for FM-15UA-OH manufactured by .

Event Text Entries

[2730500] Event description: pt nasal cannula was connected to wall-piped medical air via flow meter with yellow color coded christmas tree nipple adapter instead of connected to wall oxygen via flow meter oxygen with green color christmas tree nipple adapter.
Patient Sequence No: 1, Text Type: D, B5


[9921479] Amvex qa personnel ((b)(4)) spoke to initial reporter (b)(6) on (b)(4) at 3:00 pm regarding further explanation of incident as reported on medwatch report (b)(4). Based on further info, the flowmeter in question functioned as per our design and operational specifications. Medical staff did the initial hook up of the cannula and as described there was an error made by medical staff. The hook up of the nasal cannula was made to the medical air flowmeter rather than to the oxygen flowmeter as required by pt. There was no problem reported prior to or after the incident regarding the operation of either the oxygen or medical air flowmeters. Note: no model number, lot number or serial number were provided by initial reporter.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9617620-2012-00005
MDR Report Key2621219
Report Source00
Date Received2012-06-12
Date of Report2012-04-26
Date of Event2012-04-17
Report Date2012-04-01
Date Reported to FDA2012-04-01
Date Mfgr Received2012-05-07
Date Added to Maude2012-10-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street1111 LAKESIDE DR.
Manufacturer CityGURNEE IL 60031
Manufacturer CountryUS
Manufacturer Postal60031
Manufacturer Phone8478550500
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Generic NameFLOWMETER, OXYGEN
Product CodeCAX
Date Received2012-06-12
Catalog NumberFM-15UA-OH
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-06-12