ADVIA CENTAUR XP PROGESTERONE (PRGE) N/A 02382928

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-06-19 for ADVIA CENTAUR XP PROGESTERONE (PRGE) N/A 02382928 manufactured by Siemens Healthcare Diagnostics, Inc..

Event Text Entries

[2763895] A false low advia centaur xp progesterone result was obtained by the customer on a patient sample and the result was questioned by the physician. The customer performed repeat testing on the discordant sample and the result was higher and aligned with the patient's clinical picture. There was no report of adverse health consequences due to the discordant advia centaur xp progesterone result.
Patient Sequence No: 1, Text Type: D, B5


[9919883] The cause for the initially discordant low result when compared to the patient's clinical picture and the repeat test result is unknown. The customer's quality control results were within acceptable limits and there were no other known discordant progesterone patient results reported. No conclusion can be drawn.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1219913-2012-00245
MDR Report Key2621689
Report Source05,06
Date Received2012-06-19
Date of Report2012-05-24
Date of Event2012-05-21
Date Mfgr Received2012-05-24
Date Added to Maude2012-10-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. STEPHEN PERRY
Manufacturer Street333 CONEY ST.
Manufacturer CityEAST WALPOLE MA 02032
Manufacturer CountryUS
Manufacturer Postal02032
Manufacturer Phone5086604163
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Manufacturer Street333 CONEY STREET
Manufacturer CityEAST WALPOLE MA 02032
Manufacturer CountryUS
Manufacturer Postal Code02032
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADVIA CENTAUR XP PROGESTERONE (PRGE)
Generic NamePROGESTERONE
Product CodeJLS
Date Received2012-06-19
Model NumberN/A
Catalog Number02382928
Lot Number258
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS, INC.
Manufacturer AddressTARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2012-06-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.