VITROS 5,1 FS CHEMISTRY SYSTEM 6801375

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,05 report with the FDA on 2012-06-26 for VITROS 5,1 FS CHEMISTRY SYSTEM 6801375 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[2728853] The customer obtained discordant vitros phyt results for three patient samples (sample 1 result <3. 0 ug/ml; sample 2 result = 20. 8 ug/ml; sample 3 result = 5. 3 ug/ml) while using the vitros 5,1 fs chemistry system. Biased results of the direction and magnitude observed may lead to inappropriate physician action. The affected phyt patient results were reported to a clinician, who questioned the results. The affected patient samples were repeated at an external laboratory (repeat sample 1 result = 5. 1 ug/ml; repeat sample 2 result = 17. 0 ug/ml; repeat sample 3 result = 11. 1 ug/ml), and the repeat results were considered to be correct. However, there was no change in patient treatment based on the initial report results. There were no allegations of harm to patients as a result of this event. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[10114190] The investigation confirmed that unexpected vitros phyt patient results were obtained while using the vitros 5,1 fs analyzer. Precision testing indicated that the instrument was not performing as expected at the time of the event. An ocd field engineer performed service actions to the incubator subsystem and performed related adjustments. Performance testing following service confirmed that the equipment was returned to expected operation. The assignable root cause is an instrument related issue. There is no evidence that the vitros phyt slides malfunctioned.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319681-2012-00125
MDR Report Key2628226
Report Source00,01,05
Date Received2012-06-26
Date of Report2012-06-25
Date of Event2012-05-25
Date Mfgr Received2012-05-25
Device Manufacturer Date2008-06-25
Date Added to Maude2012-10-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JOSEPH FALVO
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal Code14626
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS 5,1 FS CHEMISTRY SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeDKH
Date Received2012-06-26
Catalog Number6801375
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2012-06-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.