SET, EXTENSION BLOODLINE FOR CRRT 36 IN CUSTOM UM-872-36

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2012-06-11 for SET, EXTENSION BLOODLINE FOR CRRT 36 IN CUSTOM UM-872-36 manufactured by Molded Products, Inc..

Event Text Entries

[2727067] Leaking occurred in the new continuous renal replacement therapy (crrt) tubing. The leaking occurred at the connection of the calcium to the venous access line. It is leaking at a permanent connection, not the luer lock. The tubing was saved. The blood was returned and new filter was set up. The tubing began leaking after the patient was repositioned in bed. The practitioner might have been able to only change out the extension tubing rather than the whole set, but decided to change out the entire set. The set was three days old and the tmp(transmembrane pressure) was rising. We are awaiting a response from the mfr.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2633354
MDR Report Key2633354
Date Received2012-06-11
Date of Report2012-06-11
Date of Event2012-05-07
Report Date2012-06-11
Date Reported to FDA2012-06-11
Date Added to Maude2012-06-28
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSET, EXTENSION BLOODLINE FOR CRRT 36 IN CUSTOM
Generic NameDIALYSIS EQUIPMENT (DESCRIBE BELOW)
Product CodeFID
Date Received2012-06-11
Returned To Mfg2012-05-14
Model NumberUM-872-36
Catalog NumberUM-872-36
Lot NumberPOTENTIALLY LOT # 80106.
ID Number*
Device AvailabilityR
Device Age0 DY
Device Sequence No1
Device Event Key0
ManufacturerMOLDED PRODUCTS, INC.
Manufacturer Address1112 CHATBURN AVENUE HARLAN IA 51537 US 51537


Patients

Patient NumberTreatmentOutcomeDate
10 2012-06-11