[19225472] Doctor reported a study pt presented with redness, irritation, and foreign body sensation of the left eye. Examination revealed corneal infiltrates os>od and pt was treated with vigamox in the left eye. Lens wear was discontinued and pt was exited from the study. Pt returned for a post-study visit 9 days later and condition had resolved. Doctor noted a small (<1mm) non-visually significant anterior stromal scar that was within the central 4mm of the left cornea. The scar was not present prior to the study. Doctor was unable to determine the exact etiology as both the test lens and the solution were new to the pt. A culture was not performed as there was no suspicion of infection.
Patient Sequence No: 1, Text Type: D, B5


[19474388] The product was returned for eval and results found the solution met chemical specs. There are no other complaints on the lot. Doctor was unable to determine the exact etiology as both the test lens and the solution were new to the pt. Based on all info, no causal factor can be determined and no conclusion can be drawn.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1313525-2012-00016
MDR Report Key2633797
Report Source02,05
Date Received2012-06-21
Date of Report2012-05-21
Date of Event2012-05-21
Date Mfgr Received2012-05-30
Device Manufacturer Date2012-02-01
Date Added to Maude2012-07-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARLENE MARTRANO
Manufacturer Street1400 NORTH GOODMAN ST.
Manufacturer CityROCHESTER NY 14609
Manufacturer CountryUS
Manufacturer Postal14609
Manufacturer Phone5853388995
Manufacturer G1BAUSCH + LOMB
Manufacturer Street5807 PELHAM ROAD
Manufacturer CityGREENVILLE SC 29615
Manufacturer CountryUS
Manufacturer Postal Code29615
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRENU FRESH MULTI-PURPOSE SOLUTION
Product CodeLYL
Date Received2012-06-21
Returned To Mfg2012-06-01
Lot NumberGB2026
Device Expiration Date2014-02-28
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAUSCH + LOMB
Manufacturer AddressROCHESTER NY US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2012-06-21

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.