OLYMPUS TELESCOPE A22002A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2012-06-22 for OLYMPUS TELESCOPE A22002A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[2705689] The user facility reported that at the start of the transurethral resection of prostate (turp) procedure, the image was very good, but after ten minutes in the procedure, the user experienced complete loss of image on the monitor. The user withdrew the telescope to be cleaned and then reinserted the telescope to the pt, but there was still no image on the monitor. The procedure was completed using a monopolar system. There was no pt injury reported.
Patient Sequence No: 1, Text Type: D, B5


[9972323] Olympus followed up with the user facility to obtain add'l info regarding the report, but with no result. The subject device was returned to olympus for evaluation. The evaluation revealed that the image was dark and blurry, which impaired good visualization. The image difficulty was attributed to a bent outer tube and broken optical system lens. The device will be serviced and returned to the user facility. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2012-00032
MDR Report Key2639505
Report Source01,08
Date Received2012-06-22
Date of Report2012-05-25
Date Mfgr Received2012-05-25
Date Added to Maude2012-10-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS TELESCOPE
Generic NameTELESCOPE
Product CodeFBP
Date Received2012-06-22
Returned To Mfg2012-06-01
Model NumberA22002A
Catalog NumberA22002A
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG GM


Patients

Patient NumberTreatmentOutcomeDate
10 2012-06-22

© 2026 FDA.report