I-CAT 0.830.4000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2012-07-02 for I-CAT 0.830.4000 manufactured by Imaging Sciences International, Llc.

Event Text Entries

[15142970] "the operator took the scan and just before the reconstruction completed, they got a "not enough memory" message. They pressed ok and the image came up. They did a workup and created a report, which they sent out. The surgeon rang them back pointing out that the image had been reversed. I have attached images. When they went to burn a cd they noticed there was no ct data, only raw. We then reconstructed and the pan was still reversed, but then it reverted to correct position a few moments later. "
Patient Sequence No: 1, Text Type: D, B5


[15277662] The configuration consists of the i-cat imaging unit, pc workstation running visionq 1. 8. 1 imaging software connected in a pacs networked environment. The investigation determined that the raw ct data was present, but the ct image was missing in the study list generated by the software. The incident was reported to have been preceded by an "out of memory" error. The engineers investigating the "out of memory" error and reversed image were not able to duplicate the error/condition in the lab, we are therefore unable to confirm the complaint. No pt injury is associated with this complaint.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2530069-2012-00011
MDR Report Key2645018
Report Source08
Date Received2012-07-02
Date of Report2012-06-27
Date of Event2012-06-04
Date Mfgr Received2012-06-04
Device Manufacturer Date2009-12-11
Date Added to Maude2012-10-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELIZABETH LAZARO
Manufacturer Street1910 NORTH PENN RD.
Manufacturer CityHATFIELD PA 19440
Manufacturer CountryUS
Manufacturer Postal19440
Manufacturer Phone2679540365
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameI-CAT
Generic NameSYSTEM, X-RAY, EXTRAORAL SOURCE, DIGITAL
Product CodeMUH
Date Received2012-07-02
Catalog Number0.830.4000
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerIMAGING SCIENCES INTERNATIONAL, LLC
Manufacturer Address1910 NORTH PENN RD. HATFIELD PA 19440 US 19440


Patients

Patient NumberTreatmentOutcomeDate
10 2012-07-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.