FRESENIUS DIALYSIS MACHINE 2008H

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2009-05-18 for FRESENIUS DIALYSIS MACHINE 2008H manufactured by Fresenius Usa.

Event Text Entries

[18850435] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 1, Text Type: D, B5


[19040341] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[18893649] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 2, Text Type: D, B5


[19153703] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 2, Text Type: N, H10


[19023880] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 3, Text Type: D, B5


[19252676] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 3, Text Type: N, H10


[18895578] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 4, Text Type: D, B5


[19154564] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 4, Text Type: N, H10


[18884955] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 5, Text Type: D, B5


[19154565] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 5, Text Type: N, H10


[19004989] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 6, Text Type: D, B5


[19252680] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 6, Text Type: N, H10


[19004990] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 7, Text Type: D, B5


[19252681] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 7, Text Type: N, H10


[18885784] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 8, Text Type: D, B5


[19154568] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 8, Text Type: N, H10


[19005929] During a dialysis conversion, we discovered on (b)(6) 2009, that the incorrect concentrate family had been chosen when the 2008h machines were re-calibrated, but the machines were running within normal limits. There existed a potential to have exposed patients to incorrect dialysate levels. We reviewed all labs and hospitalizations and found the following events may have been caused or exacerbated by the incorrectly calibrated machines: all patients admitted for fluid overload. Hospitalized on (b)(6) 2008; discharged (b)(6) 2009. On (b)(6) 2009, immediately upon discovery of the incorrect calibrations, the h machines were corrected, causing minor interruptions in treatment. After the machines were corrected, all the machines in the clinic were double checked by the chief technical officer. Both technical and clinical staff were re-educated.
Patient Sequence No: 9, Text Type: D, B5


[19252682] (b)(4) - addressed in device labeling as serious injury, addressed in device labeling as hypervolemia. (b)(4) - incorrect setup - addressed in device labeling. Corrective action was implemented at the facility prior to the manufacturer's device investigation. A fresenius equipment specialist verified that the implemented corrective action on the machines was done correctly. The machines are now set correctly for the type of concentrate being used. The facility converted from one acid concentrate type (9000 series or 36. 83x) to another type (4000 series or 45x). An equipment technician improperly set the 2008h machine for the respective concentrate type. This operator error prevented the machine from properly proportioning the acid and bicarbonate concentrate. The machine proportions acid and bicarbonate to the final dialysate concentration, based on a defined dilution ratio for the given concentrate type (36. 83x versus 45x). In house testing, by the manufacturer, simulating the operator error showed that improperly setting the machine to another concentrate type could result in a normal dilution of sodium and a bicarbonate concentration that is approximately 20% higher than the set value. The 2008h operator's manual warns the operator to "be sure to use compatible concentrate types, labeling and set-ups. These setups include machine calibration, special adapters for certain concentrate types, correct setting of concentrate option, and labeling. Failure to use the properly matched solutions and machine calibrations may allow improper dialysate to be delivered to the patient, resulting in patient injury or death. Verify composition, conductivity, and ph after converting to a different type of concentrate. " "the conversion to other dialysate concentrate types must be done by a qualified, authorized person according to written procedures. " "operator should always check conductivity and approximate ph of the dialysate prior to initiating treatment. " (b)(4).
Patient Sequence No: 9, Text Type: N, H10


MAUDE Entry Details

Report Number2937457-2009-00010
MDR Report Key2645291
Report Source05,06,07
Date Received2009-05-18
Date of Report2009-05-01
Date of Event2009-03-03
Date Facility Aware2009-03-02
Report Date2009-04-20
Date Reported to Mfgr2009-04-20
Date Mfgr Received2009-05-04
Device Manufacturer Date2000-01-01
Date Added to Maude2012-07-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELVY DIZON, BSN, RN, CNN
Manufacturer Street2637 SHADELANDS DR.
Manufacturer CityWALNUT CREEK CA 94598
Manufacturer CountryUS
Manufacturer Postal94598
Manufacturer Phone9252950200
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFRESENIUS DIALYSIS MACHINE
Generic Name2008H
Product CodeFKP
Date Received2009-05-18
Model Number2008H
OperatorOTHER
Device AvailabilityY
Device Age9 YR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerFRESENIUS USA
Manufacturer Address2637 SHADELANDS DR. WALNUT CREEK CA 94598 US 94598


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2009-05-18
201. Hospitalization 2009-05-18
301. Hospitalization 2009-05-18
401. Hospitalization 2009-05-18
501. Hospitalization 2009-05-18
601. Hospitalization 2009-05-18
701. Hospitalization 2009-05-18
801. Hospitalization 2009-05-18
901. Hospitalization 2009-05-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.