WATER TRAP AD W/TWT VLV NS 50/CS 5275P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2012-07-25 for WATER TRAP AD W/TWT VLV NS 50/CS 5275P manufactured by Carefusion.

Event Text Entries

[19222962] The customer reported that "when using the 5276p with the 0p1775 circuit and a heated humidifier, after time and the circuit has heated up, the water trap will disconnect from the circuit. " they state that the connection is slick and loose.
Patient Sequence No: 1, Text Type: D, B5


[19447894] (b)(4): results of investigation: there was no sample available for evaluation. However, an attempt to duplicate the failure reported with tubing (b)(4) to compare the suction force with the water trap 5275p according to the pull test results did not show a significant variance between the assemblies using both tubing. Therefore, the complaint was not confirmed. The device history record for the lot reported was evaluated for any issues related with this complaint. The product was manufactured, inspected and released in accordance with internal procedures and no issues were observed. The lot number reported by customer was reviewed and no anomalies were found. Our manufacturing process was reviewed and no problems were found related with the issue reported. Complaint identified for submission as an mdr following a retrospective review of (b)(4) self-manufactured complaints under capa (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030673-2012-00194
MDR Report Key2669383
Report Source00
Date Received2012-07-25
Date of Report2010-03-04
Date of Event2010-03-04
Date Mfgr Received2010-03-04
Date Added to Maude2012-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJILL RITTORNO
Manufacturer Street1500 WAUKEGAN ROAD
Manufacturer CityMCGAW PARK IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8474737294
Manufacturer G1PRODUCTOS UROLOGOS DE MEXICO S.A. DE C.V
Manufacturer StreetCERRADA VIA DE LA PRODUCCION NO85PARQUE INDUST.MEXICALI III
Manufacturer CityMEXICALI 21600
Manufacturer CountryMX
Manufacturer Postal Code21600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWATER TRAP AD W/TWT VLV NS 50/CS
Generic NameDRAIN, TEE (WATER TRAP)
Product CodeBYH
Date Received2012-07-25
Model Number5275P
Lot NumberY09H1067
OperatorOTHER
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address1500 WAUKEGAN RD MCGAW PARK IL 60085 US 60085


Patients

Patient NumberTreatmentOutcomeDate
10 2012-07-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.