REAGENT RED BLOOD CELLS BIOTESTCELL 3 816085100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2010-12-15 for REAGENT RED BLOOD CELLS BIOTESTCELL 3 816085100 manufactured by Bio-rad Medical Diagnostics Gmbh.

Event Text Entries

[2779096] The customer complained that an anti-e didn't react positive with biotestcell p3 art (b)(4), lot 7039021 on tango. The customer has sent us the affected sample. The complaint product biotestcell p3 was not sent. Therefore the quality control laboratory tested the retention sample of the complained lot biotestcell p3 in combination with the sample on tango. We could confirm a negative reaction of the sample with cell 2 of biotestcell p3. The used positive and negative control reacted correctly and clearly positive respectively negative. Further testing in the diamed gel system showed negative reactions in the liss/ coombs technique. The three phase testing in the tube technique showed negative reactions at room temperature and at 37 degrees celsius. In the indirect anti-human globulin test only a +/- reaction was observed. The enzyme test with bromelain showed a 1+ positive reaction with the e positive cell. On tango the e positive enzyme treated red cell reacted strongly positive. The tests carried out show that this antibody (anti-e) is a very weak reacting antibody and preferentially reacting in the enzyme technique.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610824-2010-00017
MDR Report Key2680362
Report Source05
Date Received2010-12-15
Date of Report2010-12-09
Date of Event2010-11-09
Date Mfgr Received2010-11-12
Date Added to Maude2012-08-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDR. MARC GORZELLIK
Manufacturer StreetINDUSTRIESTRASSE 1
Manufacturer CityDREIEICH 63303
Manufacturer CountryGM
Manufacturer Postal63303
Manufacturer Phone103801528
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameREAGENT RED BLOOD CELLS BIOTESTCELL 3
Generic NameBIOTESTCELL 3
Product CodeLKJ
Date Received2010-12-15
Catalog Number816085100
Lot Number7039021
Device Expiration Date2010-11-16
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIO-RAD MEDICAL DIAGNOSTICS GMBH
Manufacturer AddressINDUSTRIESTRASSE 1 DREIEICH D-63303 GM D-63303


Patients

Patient NumberTreatmentOutcomeDate
10 2010-12-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.