OLYMPUS URETERO-RENO VIDEOSCOPE URF-V

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2012-08-01 for OLYMPUS URETERO-RENO VIDEOSCOPE URF-V manufactured by Olympus Medical Systems Corporation.

Event Text Entries

[2866553] Olympus was informed that the images went completely out. The image flickered during the first procedure but went completely out during the second procedure. There was no pt injury reported.
Patient Sequence No: 1, Text Type: D, B5


[10122446] Olympus followed-up with the user facility to obtain add'l info regarding this report, but no further detail was provided. The device referenced in this report was returned to olympus for eval. The eval confirmed the user's report. The image was flickering intermittently and blacked out due to fluid invasion. The control body was noted with corrosion and fluid invasion. The switch buttons 1-4 were not functioning and there was a cut noted on the bending section. There was a leakage observed from the bending cover. Additionally, the internal angulation wires were damaged. The reported phenomena were attributed to mishandling.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010047-2012-00242
MDR Report Key2700475
Report Source06,07
Date Received2012-08-01
Date of Report2012-07-03
Date Mfgr Received2012-07-03
Date Added to Maude2012-09-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO 192-8507
Manufacturer CountryJA
Manufacturer Postal Code192-8507
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS URETERO-RENO VIDEOSCOPE
Generic NameURETEROSCOPE
Product CodeGCQ
Date Received2012-08-01
Returned To Mfg2012-07-06
Model NumberURF-V
Catalog NumberURF-V
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2012-08-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.