TERUMO CDI 100 MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-08-21 for TERUMO CDI 100 MONITOR manufactured by Terumo Cardiovascular Systems Corp..

Event Text Entries

[2761138] It was reported that the numbers displayed on the monitor are no longer accurate. No other details regarding the nature of this event were provided.
Patient Sequence No: 1, Text Type: D, B5


[10030480] Lab analysis found the unit failed the color chip test and has a defective opto probe related to the age of the device. No parts are available for repair. If additional info becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1828100-2012-01100
MDR Report Key2711116
Report Source05,06
Date Received2012-08-21
Date of Report2012-07-26
Date of Event2012-07-26
Date Mfgr Received2012-07-26
Date Added to Maude2012-12-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKIPP DURBIN
Manufacturer Street6200 JACKSON RD.
Manufacturer CityANN ARBOR MI 48103
Manufacturer CountryUS
Manufacturer Postal48103
Manufacturer Phone7346634145
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTERUMO CDI 100 MONITOR
Product CodeDTY
Date Received2012-08-21
Returned To Mfg2012-07-31
Model Number100
Catalog Number100
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTERUMO CARDIOVASCULAR SYSTEMS CORP.
Manufacturer AddressANN ARBOR MI 48103 US 48103


Patients

Patient NumberTreatmentOutcomeDate
10 2012-08-21

© 2026 FDA.report