VITROS 350 CHEMISTRY SYSTEM 6802153

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2012-08-24 for VITROS 350 CHEMISTRY SYSTEM 6802153 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[2799547] The customer obtained a non-reproducible, higher than expected vitros phyt proficiency sample result (api# ch-08 result = 18. 8 ug/ml vs. And expected result of 14. 53 ug/ml) while using the vitros 350 chemistry system. Biased results of the magnitude and direction observed may lead to inappropriate physician action if the event were to recur undetected with patient samples. Patient samples were not known to have been affected. There was no allegation of harm to patients as a result of this event. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[10012580] The investigation confirmed that a non-reproducible, higher than expected vitros phyt proficiency sample result was obtained while using the vitros 350 analyzer. An ocd field engineer performed service actions to multiple analyzer subsystems. Performance testing following service verified that the equipment was returned to expected operation. The investigation determined that the most likely assignable cause of the event is an instrument related issue. There is no evidence that the vitros phyt slides had malfunctioned.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319681-2012-00168
MDR Report Key2714110
Report Source05
Date Received2012-08-24
Date of Report2012-08-24
Date of Event2012-07-25
Date Mfgr Received2012-07-25
Device Manufacturer Date2012-05-24
Date Added to Maude2012-12-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JOSEPH FALVO
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal Code14626
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS 350 CHEMISTRY SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeDKH
Date Received2012-08-24
Catalog Number6802153
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2012-08-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.