OLYMPIC COOL-CAP SYSTEM 60010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-09-05 for OLYMPIC COOL-CAP SYSTEM 60010 manufactured by Natus Medical Incorporated.

Event Text Entries

[2894016] (b)(6) being treated for neonatal hypoxic-ischemic encephalopathy with the olympic cool-cap system. Approximately 29. 5 hours into treatment, three popping sounds were heard and the control module turned off. The infant was moved to a cooling blanket while an alternate cool-cap system was brought in from another facility to complete the treatment. The infant's temperature was maintained in the target range throughout the 72-hour treatment. There was no patient or user injury.
Patient Sequence No: 1, Text Type: D, B5


[10042872] The cause of the device malfunction was a failed power supply. The power supply was scheduled to be replaced on-site as part of a field corrective action ((b)(4)) but the failure occurred before the correction could be made. The power supply will now be replaced by (b)(4) and the repaired device returned to the customer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2012-00004
MDR Report Key2742082
Report Source05,06
Date Received2012-09-05
Date of Report2012-09-05
Date of Event2012-08-10
Date Mfgr Received2012-08-10
Device Manufacturer Date2008-03-01
Date Added to Maude2012-12-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685170
Single Use3
Previous Use Code3
Removal Correction Number3018859-05/25/12-002-C
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP SYSTEM
Generic NameCOOLING CAP (INFANTS)
Product CodeMXM
Date Received2012-09-05
Returned To Mfg2012-08-31
Model Number60010
Catalog Number60010
Lot NumberNOT APPLICABLE
ID NumberVERSION 2.0 SOFTWR
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer AddressSEATTLE WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-05

© 2026 FDA.report