MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-09-05 for CHOLESTECH LDX 14-697 manufactured by Alere San Diego, Inc.

Event Text Entries

[16536730] Customer alleges sparks coming from one of two frayed power cords. Customer has more than one cholestech meter and cannot identify which power cord sparked.
Patient Sequence No: 1, Text Type: D, B5


[16667848] Both power supply cables were damaged on the dc side by the converter box. Sparking was observed on one of the power supplies, when the cable with the damage was moved around. No other issue was observed during investigation. The power cord of the ldx, under extreme conditions of bending and manipulation, may eventually cut thereby exposing the copper wire being insulated. When the insulation breaks, it does so beside the strain relief of the dc cord. The ldx power cable for 9v dc was upgraded to address the cable breaking after the repeated bending. The risk associated was found to be low. Further investigation was performed in (b)(4) 2012. The risk associated with the sparking power cable was determined to be low. No further investigation will be pursued under this case number. The ldx power cable for 9v dc was upgraded to address the cable breaking after the repeated bending. The risk associated was found to be low. Corrective action not required at this time.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2012-01279
MDR Report Key2742282
Report Source05,06
Date Received2012-09-05
Date of Report2012-09-05
Date of Event2012-07-31
Date Mfgr Received2012-07-06
Date Added to Maude2012-12-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCARMEN BERGELIN, MGR
Manufacturer Street9975 SUMMERS RIDGE RD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052256
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHOLESTECH LDX
Generic NameCHOLESTEROL TEST
Product CodeCHH
Date Received2012-09-05
Returned To Mfg2012-08-16
Model Number14-697
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.