OLYMPUS TELESCOPE A50021A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-09-05 for OLYMPUS TELESCOPE A50021A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[18415719] The user facility reported that during a therapeutic laparoscopic cholecystectomy, the image was lost. The videoscope was withdrawn from the pt and the procedure was completed using a different but similar device. There was no pt injury reported.
Patient Sequence No: 1, Text Type: D, B5


[18677856] The device was returned to olympus for eval. The eval was not able to confirm the user's report. However, the user facility returned a picture captured during the use of the subject device; the picture appeared smeared and with no visualization. This type of image issue is indicative of fluid on the video connector. There were minor dents noted on the outer tube and on the distal end of the video connector. The light guide fiber on the light guide cable was sunk, the video cable was buckled and the light transmission test failed, the image difficulty was attributed to a damaged light guide cable. The subject device was serviced and returned to the user facility. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2012-00102
MDR Report Key2748113
Report Source06
Date Received2012-09-05
Date of Report2012-08-17
Date of Event2012-08-17
Date Mfgr Received2012-08-17
Date Added to Maude2012-10-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRABE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS TELESCOPE
Generic NameTELESCOPE
Product CodeFBP
Date Received2012-09-05
Returned To Mfg2012-08-20
Model NumberA50021A
Catalog NumberA50021A
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRABE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.