Event Text Entries

[2882594] The user facility reported that during a laparoscopic appendectomy, the user experienced a total loss of image. The procedure was completed with a different, but similar device. There was no pt injury.
Patient Sequence No: 1, Text Type: D, B5


[10180803] The device referenced in this report was returned to olympus for eval. The eval was able to duplicate the user's report of image loss. The image was completely black with some flickering. The eval noted minor debris underneath the light guide cover glass. The outer tube was critically bent and the device did not pass the light transmission testing. There were dents and scratches on the video connector. The image difficulty was attributed to a damaged video cable. The device was serviced and returned to the user facility. The report is being submitted as a medical device report in abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2012-00101
MDR Report Key2748126
Report Source06
Date Received2012-09-05
Date of Report2012-08-13
Date of Event2012-08-11
Date Mfgr Received2012-08-13
Date Added to Maude2012-10-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORM-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRABE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VIDEO TELESCOPE
Generic NameVIDEO TELESCOPE
Product CodeFBP
Date Received2012-09-05
Returned To Mfg2012-08-14
Model NumberA50021A
Catalog NumberA50021A
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRABE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.