3M STERI-VAC 5XL STERILIZER/AERATOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,06,07 report with the FDA on 2012-09-14 for 3M STERI-VAC 5XL STERILIZER/AERATOR manufactured by 3m Health Care.

Event Text Entries

[2891919] 3m received information from a biomedical engineer that reported he was exposed to ethylene oxide (eo) when he was repairing a 3m steri-vac 5xl sterilizer that had stopped functioning and was displaying an e73 error. Reportedly, he was not wearing a monitor and there was no alarm in the room. He stated he tried to perform a "clear silent abort" as instructed on page 45 of the service manual but the sterilizer didn't respond. He then terminated the cycle by turning the power off, set the dip switches to invoke the "reset cycle to idle" process, turned the power back on, manually opened the door (while gas was in the chamber) and pressed the stop button (which has no effect after the reset cycle to idle procedure has been performed). The reported exposure occurred because, although the engineer did check the error code listing on page 45, he did not go to page 76 for the proper and safe address of the e73 error code which says do not cycle the power.
Patient Sequence No: 1, Text Type: D, B5


[10184875] No other adverse pt effects were reported. If additional information is received, a follow-up report will be submitted. No evaluation will be performed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2110898-2012-00035
MDR Report Key2750059
Report Source04,06,07
Date Received2012-09-14
Date of Report2012-06-21
Date of Event2012-06-06
Date Mfgr Received2012-06-21
Date Added to Maude2012-09-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANDREA GRABANSKI
Manufacturer Street3M CENTER, BLDG. 275-5W-06
Manufacturer CityST. PAUL MN 55144
Manufacturer CountryUS
Manufacturer Postal55144
Manufacturer Phone6517331047
Manufacturer G1BALCHEM
Manufacturer Street95 BALCHEM LANE
Manufacturer CityGREEN POND SC 29446544
Manufacturer CountryUS
Manufacturer Postal Code29446 5443
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M STERI-VAC 5XL STERILIZER/AERATOR
Generic Name880.6860 STERILIZER, ETHYLENE-OXIDE GAS
Product CodeFLF
Date Received2012-09-14
Model NumberNA
Catalog Number5XL
Lot NumberNA
ID NumberUPC# 00707387523749
OperatorOTHER
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer AddressST. PAUL MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-14

© 2026 FDA.report