HAMMER 700 GRAMS 399,43

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2012-09-19 for HAMMER 700 GRAMS 399,43 manufactured by Synthes Gmbh.

Event Text Entries

[2906463] During a dhs procedure, the surgeon gave a gentle tap to the mallet and it fell apart. The surgeon used another instrument to complete the procedure. There was no adverse event to the pt.
Patient Sequence No: 1, Text Type: D, B5


[10152009] Device is an instrument and is not implanted/explanted. The investigation could not be completed; no conclusion could be drawn, as the product is entering the complaint system. Synthes (b)(4) was not able to complete the device history records review because the part is approx 8 years old and cannot be found in the archive system.
Patient Sequence No: 1, Text Type: N, H10


[10402787] The returned hammer was received in 2 pieces. The handle with shaft and the hammer head. The hammer head has the dowel pin still intact and secured inside the hammer head and the very tip portion of the shaft trapped by the dowel pin. The shear occurred to the shaft at the midpoint of the interface with the dowel pin. The shaft material at the shear point is homogeneous. Several dents and scratches exist on the hammer head. The returned hammer is almost 8 years old and shows noticeable wear from repeated use. The low complaint rate for this complaint issue confirms that the design is adequate for the intended use.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030965-2012-00867
MDR Report Key2759371
Report Source05,07
Date Received2012-09-19
Date of Report2012-08-22
Date of Event2012-08-22
Date Mfgr Received2012-10-10
Device Manufacturer Date2004-11-29
Date Added to Maude2013-01-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSARMA PIN
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Manufacturer G1SYNTHES GMBH
Manufacturer StreetEIMATTSTRASSE 3
Manufacturer CityOBERDORF CH-4436
Manufacturer CountrySZ
Manufacturer Postal CodeCH-4436
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHAMMER 700 GRAMS
Generic NameHAMMER 700 GRAMS
Product CodeHXL
Date Received2012-09-19
Returned To Mfg2012-09-07
Catalog Number399,43
Lot NumberA7NA44
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES GMBH
Manufacturer AddressOBERDORF SZ


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.