[2907904] The customer reported that the computer would not boot up and the power supply emitted smoke when the computer was turned on.
Patient Sequence No: 1, Text Type: D, B5


[10136003] Information in section f10 completed by the manufacturer. The field service engineer inspected the system at the customer location and observed that the computer''s power supply had a burnt smell and the video board had a faulty capacitor. The field service engineer replaced the system computer and verified the system operation following repair.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006695864-2012-00157
MDR Report Key2763263
Report Source06,07
Date Received2012-09-27
Date of Report2012-08-29
Date of Event2012-08-29
Date Mfgr Received2012-08-29
Device Manufacturer Date2011-05-05
Date Added to Maude2012-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCAROL KAIL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478985
Manufacturer G1ABBOTT MEDICAL OPTICS INC.
Manufacturer StreetAMO MANUFACTURING 510 COTTONWOOD
Manufacturer CityMILPITAS CA 95035
Manufacturer CountryUS
Manufacturer Postal Code95035
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWAVESCAN ADVANCED CUSTOMVUE
Generic NameREFRACTIVE MEASUREMENT
Product CodeHKO
Date Received2012-09-27
Model NumberACV
Catalog Number0070-1478
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MEDICAL OPTICS
Manufacturer AddressSANTA ANA CA


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.