MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2012-09-27 for WAVESCAN ADVANCED CUSTOMVUE ACV 0070-1478 manufactured by Abbott Medical Optics.
[2907904]
The customer reported that the computer would not boot up and the power supply emitted smoke when the computer was turned on.
Patient Sequence No: 1, Text Type: D, B5
[10136003]
Information in section f10 completed by the manufacturer. The field service engineer inspected the system at the customer location and observed that the computer''s power supply had a burnt smell and the video board had a faulty capacitor. The field service engineer replaced the system computer and verified the system operation following repair.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3006695864-2012-00157 |
MDR Report Key | 2763263 |
Report Source | 06,07 |
Date Received | 2012-09-27 |
Date of Report | 2012-08-29 |
Date of Event | 2012-08-29 |
Date Mfgr Received | 2012-08-29 |
Device Manufacturer Date | 2011-05-05 |
Date Added to Maude | 2012-09-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | CAROL KAIL |
Manufacturer Street | 1700 EAST ST. ANDREW PLACE |
Manufacturer City | SANTA ANA CA 92705 |
Manufacturer Country | US |
Manufacturer Postal | 92705 |
Manufacturer Phone | 7142478985 |
Manufacturer G1 | ABBOTT MEDICAL OPTICS INC. |
Manufacturer Street | AMO MANUFACTURING 510 COTTONWOOD |
Manufacturer City | MILPITAS CA 95035 |
Manufacturer Country | US |
Manufacturer Postal Code | 95035 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | WAVESCAN ADVANCED CUSTOMVUE |
Generic Name | REFRACTIVE MEASUREMENT |
Product Code | HKO |
Date Received | 2012-09-27 |
Model Number | ACV |
Catalog Number | 0070-1478 |
Operator | OTHER HEALTH CARE PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MEDICAL OPTICS |
Manufacturer Address | SANTA ANA CA |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2012-09-27 |