J-TIP SYRINGE J-TIP 0.25 ML 8164-2001-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-09-19 for J-TIP SYRINGE J-TIP 0.25 ML 8164-2001-25 manufactured by National Medical Products, Inc..

Event Text Entries

[16049547] User reported malfunction, breakage of device(s) at user one unit tip broke, one unit cracked, on unit where the filling unit attaches broke off. Not normal practice but cannot report that the devices were not stored outside labeling storage conditions. Syringe broke apart during activation.
Patient Sequence No: 1, Text Type: D, B5


[16325541] One hundred percent inspection visual 100x magnification of filled devices returned from pharmacy; 76 units - nine units showed cracking - 152 units in original sterile packaging - opened 100% inspection visual 100x magnification - zero defects found - all units filled and inspected at zero time - zero defects found. Testing outside normal storage condition in process - will report following completion of testing. Previous lot testing could only replicate this type of cracking when units were filled and frozen.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027063-2012-00001
MDR Report Key2765308
Report Source06
Date Received2012-09-19
Date of Report2012-08-19
Date Mfgr Received2012-08-23
Device Manufacturer Date2012-07-01
Date Added to Maude2013-01-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street57 PARKER ST.
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9492853517
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameJ-TIP SYRINGE
Generic NameSYRINGE, NEEDLESS
Product CodeKZE
Date Received2012-09-19
Returned To Mfg2012-08-23
Model NumberJ-TIP 0.25 ML
Catalog Number8164-2001-25
Lot Number1205001
Device Expiration Date2014-05-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATIONAL MEDICAL PRODUCTS, INC.
Manufacturer Address57 PARKER ST. IRVINE CA 92718 US 92718


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-19