LEKSELL ANCHOR BOLT LSB-AN-05

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01 report with the FDA on 2012-09-28 for LEKSELL ANCHOR BOLT LSB-AN-05 manufactured by Ad-tech Medicla Instrumetn Corporation.

Event Text Entries

[18062486] Two (2) anchor bolts broke during insertion. The doctor reported that the pt was not compromised and the bolts were removed. Surgery was prolonged by two (2) hours.
Patient Sequence No: 1, Text Type: D, B5


[18423429] The doctor had a problem with the twist drill and screw driver bending which affected the placement of the bolts. Two anchor bolts breaking were the results of these issues. The doctor believes the breaking is due to the bending of the screw driver while placing the bolts. Ad-tech has determined that the drill and placement wrench were not used with a guide tube to properly place the bolts, thus creating an uneven torque which could cause stress on the bolt. The screwdriver was used to remove the bolts, and the doctor was unconcerned at removing the bolts as they already had access through the skull to place the electrodes. The electrodes were sutured in place, and the doctor reported that the pt was not compromised. No permanent damage to pt, no serious injury and no further surgery was required due to the malfunction. The customer was unable to provide ad-tech with the lot numbers for the anchor bolts. The decision to file this adverse event report is the result of a re-review of complaint files.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2183456-2012-00005
MDR Report Key2768567
Report Source00,01
Date Received2012-09-28
Date Mfgr Received2011-03-03
Date Added to Maude2013-01-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLISA THEAMA
Manufacturer Street1901 WILLIAM ST.
Manufacturer CityRACINE WI 53404
Manufacturer CountryUS
Manufacturer Postal53404
Manufacturer Phone2626341555
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEKSELL ANCHOR BOLT
Generic NameANCHOR BOLT
Product CodeGYC
Date Received2012-09-28
Model NumberLSB-AN-05
Catalog NumberLSB-AN-05
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAD-TECH MEDICLA INSTRUMETN CORPORATION
Manufacturer AddressRACINE WI 53404 US 53404


Patients

Patient NumberTreatmentOutcomeDate
10 2012-09-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.