ONLINE DAT BENZODIAZEPINES PLUS 04499034190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-10-02 for ONLINE DAT BENZODIAZEPINES PLUS 04499034190 manufactured by Roche Diagnostics.

Event Text Entries

[19387581] The customer provided the involved patient sample for investigation. It was determined that the sample was negative with the complained benzodiazepines reagent. For gc-ms measurement, an enzymatic glucuronidase cleavage is typically performed, so the measurement would be a sum of oxazepam, lorazepam and its glucuronides. The benzodiazepines assay that the customer used has very low cross reactivity for glucuronides (oxazepam glucuronide is not cited in the package insert, but the structure is similar to lorazepam glucuronide and the cross reactivity for this is claimed at approximately (b)(4)).
Patient Sequence No: 1, Text Type: N, H10


[20791415] The customer reported that they received erroneous results for one patient urine sample tested for benzodiazepines plus (benzodiazepines). The customer uses a semiquantitative application of the assay with a 200 ng/ml cutoff. The sample initially recovered a value of 143 ng/ml which is considered negative. The result was reported outside of the laboratory. The customer received a complaint from their client, so the sample was repeated. The sample was sent to a reference laboratory for confirmation where it was tested using gc/ms on (b)(6) 2012. The gc/ms result was negative for nordiazepam (100 ng/ml cutoff), 140 ng/ml for oxazepam (100 ng/ml cutoff), negative for temazepam (100 ng/ml cutoff), 640 ng/ml for lorazepam (100 ng/ml cutoff), negative for flurazepam metabolite (100 ng/ml cutoff), and negative for alprazolam (50 ng/ml cutoff for alpha- hydroxyalprazolam). The sample was aliquoted and both the aliquot and original sample were repeated using the same p module analyzer on (b)(6) 2012. The repeat of the aliquot resulted as 111 ng/ml and the repeat of the original sample resulted as 130 ng/ml. The sample aliquot and original sample were also repeated a second time using a different p module analyzer (serial number (b)(4)) on (b)(6) 2012. The second repeat of the aliquot on p module serial number (b)(4) resulted as 119 ng/ml and the second repeat of the original sample on p module serial number (b)(4) resulted as 126 ng/ml. The customer believed the gc/ms confirmation results to be the correct results. The patient was not adversely affected by the event. The benzodiazepines reagent lot number was 64948901 with an expiration date of 06/30/2013. The field service representative was not able to determine a cause. He performed checks on the module for rinsing of the cells, the probes and stirrers. He checked the vacuum pressures and water pressures. He checked the probe alignments and lubricated the probe shafts. He recalibrated the assay with passing results. He also ran quality controls and the results were at or near the customer's mean. He ran precision testing and the results were within guidelines.
Patient Sequence No: 1, Text Type: D, B5


[20906987] .
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2012-04920
MDR Report Key2769641
Report Source05,06
Date Received2012-10-02
Date of Report2012-11-13
Date of Event2012-09-18
Date Mfgr Received2012-09-26
Date Added to Maude2012-10-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA JENNIFER WOLFGRAM
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175217008
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameONLINE DAT BENZODIAZEPINES PLUS
Generic NameENZYME IMMUNOASSAY, BENZODIAZIPINE
Product CodeJXM
Date Received2012-10-02
Model NumberNA
Catalog Number04499034190
Lot Number64948901
ID NumberNA
Device Expiration Date2013-06-30
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2012-10-02

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.