INVISIO DIGITAL CONTROLLER WITH DVI IDC-1500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-10-01 for INVISIO DIGITAL CONTROLLER WITH DVI IDC-1500 manufactured by Gyrus Acmi, Inc..

Event Text Entries

[2969063] It was reported that during a ureteroscopy procedure using the digital fiberoptic scope. The ccu stopped midway through the procedure and the hosp had to procedure another controller from new hampshire to complete the procedure. The pt was awakened from anesthesia as it took 1 1/2 hours to procure the new controller. The pt was placed back under anesthesia and the procedure was completed with no further problems or harm to the pt.
Patient Sequence No: 1, Text Type: D, B5


[10273249] The customer's complaint of the instrument not working was unconfirmed. The eval of the returned idc-1500 found the scope has an older software version of 4. 20. 7 installed. The scope ran connected with the test dur-d for one hour with no indication of loss of image.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004356764-2012-00001
MDR Report Key2770178
Report Source06
Date Received2012-10-01
Date of Report2012-08-29
Date of Event2012-08-29
Date Mfgr Received2012-08-29
Date Added to Maude2013-01-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTERRENCE SULLIVAN
Manufacturer Street136 TURNPIKE RD
Manufacturer CitySOUTHBOROUGH MA 01772
Manufacturer CountryUS
Manufacturer Postal01772
Manufacturer Phone5088042739
Manufacturer G1GYRUS ACMI, INC.
Manufacturer Street300 STILLWATER AVE
Manufacturer CitySTAMFORD CT 06902
Manufacturer CountryUS
Manufacturer Postal Code06902
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINVISIO DIGITAL CONTROLLER WITH DVI
Generic NameINVISIO DIGITAL CONTROLLER WITH DVI
Product CodeKQM
Date Received2012-10-01
Returned To Mfg2012-09-04
Model NumberIDC-1500
Catalog NumberIDC-1500
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI, INC.
Manufacturer Address300 STILLWATER AVE STAMFORD CT 06902 US 06902


Patients

Patient NumberTreatmentOutcomeDate
10 2012-10-01