TELESCOPE, 10MM, 0 DEGREE, HD, AUTOCLAVABLE WA53000A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2012-09-26 for TELESCOPE, 10MM, 0 DEGREE, HD, AUTOCLAVABLE WA53000A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[3035571] The physician performed a laparoscopic oophorocystectomy. During the procedure, the pt suffered a skin burn on her thigh, as the doctor had put the device in a pocket of the drape without switching the light off. The doctor carried out treatment of the burns.
Patient Sequence No: 1, Text Type: D, B5


[10192728] The device involved in the event has not been returned to mfr for investigation, but has been inspected by the local sales and service organization without observation. As the products of other manufacturers too, the videotelescope heats up at the distal end during light emission. The heating-up effect of the wa53000a is nevertheless in conformance to the requirements of (b)(4). It can be considered basic knowledge for medical professionals to switch-off the videotelescope when not in use, especially when it is put aside or on the pt. For that purpose and to meet customer expectations regarding usability, the product is equipped with a stand-by mode. Appropriate warning message are included in the ifu. The event is an obvious consequence of a use error with rather minor (temporary) consequences. The case will be closed without further actions.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2012-00083
MDR Report Key2773529
Report Source05,06,07
Date Received2012-09-26
Date of Report2012-11-16
Date of Event2011-11-16
Date Mfgr Received2012-01-13
Device Manufacturer Date2006-11-01
Date Added to Maude2012-10-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL PETRICH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal22045
Manufacturer Phone0669662572
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTELESCOPE, 10MM, 0 DEGREE, HD, AUTOCLAVABLE
Generic NameOLYMPUS VIDEOTELESCOPE
Product CodeFBP
Date Received2012-09-26
Model NumberWA53000A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG GM


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2012-09-26

© 2026 FDA.report