N LATEX ADNASE B OWTI

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-10-08 for N LATEX ADNASE B OWTI manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[18252109] A falsely depressed anti-streptococcal dnase b (adnase b) result was reported on a patient sample. The result was reported to the physician who questioned the result. The sample was re-run and a higher result was obtained in line with physician expectations. Patient treatment was not altered or prescribed on the basis of the falsely depressed adnase b result. There is no report of adverse outcome to patients as a result of falsely depressed adnase b result.
Patient Sequence No: 1, Text Type: D, B5


[18458197] The cause of the falsely depressed adnase b result was sample integrity. The n latex adnase b ifu states: serum samples must be completely coagulated and, after centrifugation, must not contain any particles or traces of fibrin. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610806-2012-00014
MDR Report Key2775448
Report Source05,06
Date Received2012-10-08
Date of Report2012-09-11
Date of Event2012-09-06
Date Mfgr Received2012-09-11
Device Manufacturer Date2012-03-22
Date Added to Maude2013-01-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Street76 EMIL VON BEHRING STRASSE
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameN LATEX ADNASE B
Generic NameN LATEX ADNASE B
Product CodeGTR
Date Received2012-10-08
Catalog NumberOWTI
Lot Number41601
Device Expiration Date2013-01-19
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Address76 EMIL VON BEHRING STRASSE MARBURG, 35041 GM 35041


Patients

Patient NumberTreatmentOutcomeDate
10 2012-10-08