OLYMPIC COOL-CAP SYSTEM 60010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2012-10-03 for OLYMPIC COOL-CAP SYSTEM 60010 manufactured by Natus Medical Inc..

Event Text Entries

[3069091] Infant being treated for neonatal hypoxic-ischemic encephalopathy with the olympic cool-cap system. The user noted a smell similar to an electrical device being overheated but the device continued to function normally. Approximately 18 hours into treatment, a click sound was heard and the control module turned off. The issue was not resolved by plugging the device into a different electrical outlet. The infant's 72-hour treatment was completed using a cool-cap system in a nearby hospital. There was no patient or user injury.
Patient Sequence No: 1, Text Type: D, B5


[10212510] Serial #: (b)(4). The cause of the device malfunction was most likely a failed power supply. The power supply was scheduled to be replaced on-site as part of a field corrective action (correction report 3018859-05/25/12-002-c; fda recall #z-1843-2012; natus capa # (b)(4)) but the failure occurred before the correction could be made. The power supply will now be replaced by natus-(b)(4) and the repaired device returned to the customer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2012-00005
MDR Report Key2776638
Report Source01,08
Date Received2012-10-03
Date of Report2012-10-03
Date of Event2012-09-05
Date Mfgr Received2012-09-05
Device Manufacturer Date2009-04-01
Date Added to Maude2013-01-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685170
Single Use3
Previous Use Code3
Removal Correction Number3018859-05/25/12-002-C
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP SYSTEM
Generic NameCOOLING CAP (INFANTS)
Product CodeMXM
Date Received2012-10-03
Model Number60010
Catalog Number60010
Lot NumberNA
ID NumberVERSION 2.0 SOFTWARE
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC.
Manufacturer AddressSEATTLE WA US