HIGH TEMPERATURE CAUTERY 6522 6522JM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2006-11-15 for HIGH TEMPERATURE CAUTERY 6522 6522JM manufactured by Medtronic Heart Valves.

Event Text Entries

[2919171] Info received indicates that the manufacture and expiration dates on the label are incorrect for one lot of product. The incorrect labeling was reported by the sales rep and there was no pt involvement.
Patient Sequence No: 1, Text Type: D, B5


[10276890] (b)(4). Analysis: the actual product was evaluated and showed the incorrect dates for manufacture and expiration. Investigation revealed that the supplier's expiration date for one lot of product was inadvertently entered as the manufacture date during re-packaging, and the expiration date was automatically set to 4 years after manufacture, therefore, making both dates incorrect. Conclusion: the event was related to labeling, and there was no pt involvement.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2127690-2006-00004
MDR Report Key2783768
Report Source01,07
Date Received2006-11-15
Date of Report2006-10-16
Date of Event2006-10-16
Date Mfgr Received2006-10-16
Date Added to Maude2012-10-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactLAURALIE MILIKAN
Manufacturer Street7611 NORTHLAND DRIVE
Manufacturer CityBROOKLYN PARK MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7633919124
Manufacturer G1MEDTRONIC HEART VALVES
Manufacturer Street8299 CENTRAL AVE., N.E.
Manufacturer CitySPRING LAKE PARK MN 55432357
Manufacturer CountryUS
Manufacturer Postal Code55432 3576
Single Use3
Remedial ActionRB
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHIGH TEMPERATURE CAUTERY
Product CodeHQP
Date Received2006-11-15
Returned To Mfg2006-11-02
Model Number6522
Catalog Number6522JM
Lot NumberB13050
ID NumberNA
Device Expiration Date2010-01-01
OperatorOTHER
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC HEART VALVES
Manufacturer Address8299 CENTRAL AVE., N.E. SPRING LAKE PARK MN 55432357 US 55432 3576


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-11-15

© 2026 FDA.report