PCP_2 REAGENT PCP_2 REAGENTS 10378764

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-10-11 for PCP_2 REAGENT PCP_2 REAGENTS 10378764 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[17971696] Seven false positive results for phencyclidine (pcp) were obtained on an advia 1800 instrument. The results were reported to the physicians and were questioned. The same samples were rerun on an alternative manual method and negative results were obtained. There are no known reports of patient intervention or adverse health consequences due to the discordant pcp results.
Patient Sequence No: 1, Text Type: D, B5


[18175385] Siemens investigated the event. The database of the instrument was investigated by the technical application specialist (tas). The cause of the discordant results was incorrect calibration. The customer changed the lot of the reagent without performing a full two-point calibration. The results of the calibration were atypical. The results of the quality control were also atypical. The instructions for use (ifu) states: " the pcp method provides only a preliminary test result. The user must use more specific alternative methods to obtain a confirmed analytical result. Siemens healthcare diagnostics recommends using the gaschromatography/mass spectroscopy (gc/ms) confirmatory method. " the instrument is performing within specifications. Further evaluation of the device is not required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2012-00333
MDR Report Key2784226
Report Source05,06
Date Received2012-10-11
Date of Report2012-10-03
Date of Event2012-05-15
Date Mfgr Received2012-10-03
Date Added to Maude2013-01-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. RANJAN SADARANGANI
Manufacturer StreetSIEMENS HEALTHCARE DIAGNOSTICS 511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242868
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePCP_2 REAGENT
Generic NameENZYME IMMUNOASSAY, PHENCYCLIDINE
Product CodeLCM
Date Received2012-10-11
Model NumberPCP_2 REAGENTS
Catalog Number10378764
OperatorHEALTH PROFESSIONAL
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2012-10-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.