RELIANCE SYNERGY WASHER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-10-16 for RELIANCE SYNERGY WASHER manufactured by Steris Canada Corporation.

Event Text Entries

[3072655] The user facility reported that water was leaking from underneath the unit. The amount of water was approximately 6 gallons. The following day the user facility reported that water was leaking from the unit. No injuries, procedural delays/cancellations or property damage reported.
Patient Sequence No: 1, Text Type: D, B5


[10246890] A steris service technician inspected the unit for the first reported event and found a small leak on the main hose and a leak to the spray arm. The technician replaced all the hoses on the recirculation system. The technician ran two test cycles and found the unit to be operating properly. The steris service technician inspected the unit the following day per the customer's report of continued leaking and could not verify where the reported water leak was coming from. He did find that the insulation on the main hose going to the spray arm to be wet, but no leak was observed. The technician verified that all hoses and clamps were secured. The technician ran two test cycles and confirmed the unit was operational; no further issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680353-2012-00086
MDR Report Key2792858
Report Source06
Date Received2012-10-16
Date of Report2012-10-16
Date of Event2012-09-20
Date Mfgr Received2012-09-20
Date Added to Maude2013-01-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS LINDSEY MCGOWAN
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927519
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRELIANCE SYNERGY WASHER
Generic NameWASHER
Product CodeLDS
Date Received2012-10-16
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2012-10-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.