WATERS QUATTRO MICRO VB-QMICRO

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2012-10-15 for WATERS QUATTRO MICRO VB-QMICRO manufactured by Waters Corp.

Event Text Entries

[3083068] During a service call, the customer reported to a waters field service engineer that on (b)(6), 2012, he had gotten four "strong static shocks" within a matter of seconds from the top cover of the quattro micro instrument that he did not think were static in nature. The customer said he was not injured and did not require medical attention.
Patient Sequence No: 1, Text Type: D, B5


[10431999] A waters field service engineer (fse) inspected the instrument at the customer's site and did not find any faults with the device. He did not find any loose wires and the instrument was appropriately grounded. During his inspection of the instrument, the fse touched the cover and chassis and did not experience any electrical discharge. No previous similar events have been reported by the customer. And a follow-up call to the customer confirmed that no similar events have occurred since this event was reported. No issues were found with the device and no injury was reported. We consider this report complete. However, if additional information is received, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616673-2012-00001
MDR Report Key2796347
Report Source00
Date Received2012-10-15
Date of Report2012-09-20
Date of Event2012-09-17
Date Mfgr Received2012-09-20
Device Manufacturer Date2003-08-01
Date Added to Maude2013-01-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL PACHECO
Manufacturer Street34 MAPLE STREET
Manufacturer CityMILFORD MA 01757
Manufacturer CountryUS
Manufacturer Postal01757
Manufacturer Phone5084822532
Manufacturer G1WATERS TECHNOLOGIES IRELAND, LTD
Manufacturer StreetWEXFORD BUSINESS PARK, DRINAGH
Manufacturer CityWEXFORD
Manufacturer CountryEI
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameWATERS QUATTRO MICRO
Generic NameDOP, MASS SPECTROMETER, CLINICAL USE
Product CodeDOP
Date Received2012-10-15
Model NumberQUATTRO MICRO
Catalog NumberVB-QMICRO
Lot NumberNA
ID NumberNA
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerWATERS CORP
Manufacturer Address34 MAPLE STREET MILFORD MA 01757 US 01757


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-10-15