VP 2000 PROCESSOR 02J11-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-01-12 for VP 2000 PROCESSOR 02J11-001 manufactured by Abbott Molecular, Inc..

Event Text Entries

[21760241] The abbott vp 2000 processor is a device designed to automate and standardize slide specimen processing and routine slide staining for the laboratory. Customer was running urovysion protocol and noticed that basin with the protease heated up to 69 degrees c when the program was set to 37 degrees c. Customer stopped the vp2000 instrument and rebooted. To troubleshoot the vp2000 customer filled the heated basins with water and started the instrument with the same protocol. The instrument heated unexpectedly and customer noticed smoke burning out of the heater module vp2000. Customer turned off the vp2000. Customer stated that the vp2000 heated basin temperatures were not being controlled properly, and that a small amount of smoke was observed near heater box #3. Field service noted that there was dried and crystallized reagent under heater box #1, and that this reagent had slowly advanced to the heater module assy. Control board ((b)(4)). Damage was observed on the heater module assembly, control board around the heater box power connectors, and it was determined that the crystallized reagent had "shorted" the heater box power connections. The circuit breaker (instrument on/off switch) was found in the "off" position, indicating that the circuit breaker tripped when the short occurred. Field service replaced the heater module assembly to resolve this issue. No injury was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005248192-2011-00001
MDR Report Key2796736
Report Source07
Date Received2011-01-12
Date of Report2011-01-12
Date of Event2010-12-17
Date Mfgr Received2010-12-17
Date Added to Maude2012-11-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR DIMITRIS DEMIRTZOGLOU
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone2243617975
Manufacturer G1ABBOTT MOLECULAR, INC.
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal Code60018
Single Use0
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVP 2000 PROCESSOR
Generic NameSTAINER, TISSUE, AUTOMATED, PRODUCT CODE: KEY
Product CodeKEY
Date Received2011-01-12
Catalog Number02J11-001
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MOLECULAR, INC.
Manufacturer Address1300 EAST TOUHY AVENUE DES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
10 2011-01-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.