BLADE TRITON 8MM GRAFT HARVEST 238GH1SS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,health professional report with the FDA on 2012-10-22 for BLADE TRITON 8MM GRAFT HARVEST 238GH1SS manufactured by Mdt Powered Surgical Solutions.

Event Text Entries

[3091183] It was reported that "there was an incident today (b)(6) where doctor asked for a 8mm and ended up taking a 6mm graft not knowing. I do not know the extent the packaging played on this but i do know there was confusion and he had to take a second graft. " no patient impact reported. No additional information was available upon follow up.
Patient Sequence No: 1, Text Type: D, B5


[10251409] Report confirmed. Visual examination of an example of the packaging determined that the description, containing the size of the blade, was not reflected on the current packaging. Gcapa (b)(4) was opened on (b)(6) 2012 to address this issue. Instructions included in device user manual states, "prior to each use, the entire triton" power surgical instrument system must be inspected for proper operation". We will continue to track and trend this complaint type.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1625507-2012-00102
MDR Report Key2798173
Report Source05,HEALTH PROFESSIONAL
Date Received2012-10-22
Date of Report2012-09-19
Date of Event2012-09-19
Date Mfgr Received2012-09-19
Date Added to Maude2012-10-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVID LEERS
Manufacturer Street4620 NORTH BEACH STREET
Manufacturer CityFT. WORTH TX 761373291
Manufacturer CountryUS
Manufacturer Postal761373291
Manufacturer Phone8177886587
Manufacturer G1MEDTRONIC POWERED SURGICAL SOLUTIONS
Manufacturer Street4620 NORTH BEACH STREET
Manufacturer CityFT. WORTH TX 761373291
Manufacturer CountryUS
Manufacturer Postal Code761373291
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBLADE TRITON 8MM GRAFT HARVEST
Generic NameSAW, POWERED, AND ACCESSORIES
Product CodeKFK
Date Received2012-10-22
Catalog Number238GH1SS
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMDT POWERED SURGICAL SOLUTIONS
Manufacturer Address4620 N BEACH ST FORT WORTH TX 76137 US 76137


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-10-22