DIMENSION VISTA? CLINICAL CHEMISTRY SYSTEM K1027

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-10-24 for DIMENSION VISTA? CLINICAL CHEMISTRY SYSTEM K1027 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[17564398] Falsely depressed cholesterol (chol) results were obtained on patient samples. Results were reported to the physician. The results were questioned within the laboratory due to occasional abnormal reaction flags on chol tests. The samples were re-run and higher results were obtained. Corrected results were reported on the patient samples. It is unknown if patient treatment was altered or prescribed. There was no report of adverse health consequences as a result of the falsely depressed chol results.
Patient Sequence No: 1, Text Type: D, B5


[17670583] Analysis of the instrument and instrument data indicate that the cause for the falsely depressed cholesterol results is incomplete hydration of the reagent cartridge well. The siemens healthcare diagnostics technical service center representative directed the customer to follow recommended prehydration of the reagent cartridge well. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2517506-2012-00282
MDR Report Key2803258
Report Source05,06
Date Received2012-10-24
Date of Report2012-09-25
Date of Event2012-09-24
Date Mfgr Received2012-09-25
Device Manufacturer Date2012-06-11
Date Added to Maude2013-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE
Manufacturer CityNEWARK DE 19702
Manufacturer CountryUS
Manufacturer Postal Code19702
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION VISTA? CLINICAL CHEMISTRY SYSTEM
Generic NameCHOLESTEROL FLEX? REAGENT CARTRIDGE
Product CodeCHH
Date Received2012-10-24
Catalog NumberK1027
Lot Number12163BB
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE NEWARK DE 19702 US 19702


Patients

Patient NumberTreatmentOutcomeDate
10 2012-10-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.