MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-10-24 for RELIANCE SYNERGY WASHER manufactured by Steris Canada Corporation.
[3026160]
The user facility reported that their reliance synergy washer/disinfector was leaking water. No procedure or injuries were reported. Damage to the linoleum floor by the water was reported.
Patient Sequence No: 1, Text Type: D, B5
[10382490]
The steris technician found that steam was leaking from the strainer cap. The technician replaced o-rings, retightened the cap, and replaced the piston for drying valve. The technician ran a test cycle and the unit was operational.
Patient Sequence No: 1, Text Type: N, H10
| Report Number | 9680353-2012-00087 |
| MDR Report Key | 2803285 |
| Report Source | 06 |
| Date Received | 2012-10-24 |
| Date of Report | 2012-10-24 |
| Date of Event | 2012-09-26 |
| Date Mfgr Received | 2012-09-26 |
| Date Added to Maude | 2013-02-04 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 0 |
| Event Location | 0 |
| Manufacturer Contact | MS LINDSEY MCGOWAN |
| Manufacturer Street | 5960 HEISLEY ROAD |
| Manufacturer City | MENTOR OH 44060 |
| Manufacturer Country | US |
| Manufacturer Postal | 44060 |
| Manufacturer Phone | 4403927519 |
| Manufacturer G1 | STERIS CANADA CORPORATION |
| Manufacturer Street | 490, ARMAND-PARIS |
| Manufacturer City | QUEBEC, QUEBEC GIC8A3 |
| Manufacturer Country | CA |
| Manufacturer Postal Code | GIC 8A3 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | RELIANCE SYNERGY WASHER |
| Generic Name | WASHER |
| Product Code | LDS |
| Date Received | 2012-10-24 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Eval'ed by Mfgr | Y |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | STERIS CANADA CORPORATION |
| Manufacturer Address | 490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2012-10-24 |