MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-10-24 for RELIANCE SYNERGY WASHER manufactured by Steris Canada Corporation.
[3026160]
The user facility reported that their reliance synergy washer/disinfector was leaking water. No procedure or injuries were reported. Damage to the linoleum floor by the water was reported.
Patient Sequence No: 1, Text Type: D, B5
[10382490]
The steris technician found that steam was leaking from the strainer cap. The technician replaced o-rings, retightened the cap, and replaced the piston for drying valve. The technician ran a test cycle and the unit was operational.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 9680353-2012-00087 |
MDR Report Key | 2803285 |
Report Source | 06 |
Date Received | 2012-10-24 |
Date of Report | 2012-10-24 |
Date of Event | 2012-09-26 |
Date Mfgr Received | 2012-09-26 |
Date Added to Maude | 2013-02-04 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MS LINDSEY MCGOWAN |
Manufacturer Street | 5960 HEISLEY ROAD |
Manufacturer City | MENTOR OH 44060 |
Manufacturer Country | US |
Manufacturer Postal | 44060 |
Manufacturer Phone | 4403927519 |
Manufacturer G1 | STERIS CANADA CORPORATION |
Manufacturer Street | 490, ARMAND-PARIS |
Manufacturer City | QUEBEC, QUEBEC GIC8A3 |
Manufacturer Country | CA |
Manufacturer Postal Code | GIC 8A3 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | RELIANCE SYNERGY WASHER |
Generic Name | WASHER |
Product Code | LDS |
Date Received | 2012-10-24 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STERIS CANADA CORPORATION |
Manufacturer Address | 490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2012-10-24 |