SEDATION SYSTEMS LLC CA 0120MX

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2012-10-12 for SEDATION SYSTEMS LLC CA 0120MX manufactured by Sedations Systems And Sedation.

Event Text Entries

[18527600] Based on a headache complaint, some nitrous oxide field screening was done on the sedation systems disposable sedation-breathing circuit and mask set. The areas identified were the assembled plastic distribution tubes. This appears to be a mfg defect in the tube materials and seemed to worsen when the tube was bent or stressed. The nitrous levels were elevated significantly above the standard for the following lots lot #981383 and lot # 89364. We suspended services, removed the product, and contacted the mfr. We requested and received qa/qc pressure test data on lot # 981383. This was a one page report that indicates a? P? For air pressure, pull test, and visual.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5027326
MDR Report Key2804116
Date Received2012-10-12
Date of Report2012-10-12
Date Added to Maude2012-10-26
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameSEDATION SYSTEMS LLC
Generic NameDISPOSABLE SEDATION-BREATHING CIRCUIT AND MASK SET
Product CodeBSJ
Date Received2012-10-12
Returned To Mfg2012-10-11
Catalog NumberCA 0120MX
Lot Number89364
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerSEDATIONS SYSTEMS AND SEDATION
Manufacturer AddressCLEARWATER FL 33759 US 33759

Device Sequence Number: 2

Brand NameSEDATION SYSTEMS LLC
Generic NameDISPOSABLE SEDATION-BREATHING CIRCUIT AND MASK SET
Product CodeBSJ
Date Received2012-10-12
Returned To Mfg2012-10-11
Catalog NumberCA 0120MX
Lot Number981383
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No2
Device Event Key0
ManufacturerSEDATIONS SYSTEMS AND SEDATION
Manufacturer AddressCLEARWATER FL 33759 US 33759


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-10-12

© 2026 FDA.report