GCE COMBILITE I VIPRS, 5 MICRO FILTER, LIV FOR OXYGEN 070137

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2008-09-22 for GCE COMBILITE I VIPRS, 5 MICRO FILTER, LIV FOR OXYGEN 070137 manufactured by Linde Gas North America Llc.

Event Text Entries

[3022081] Information was received on (b)(6) 2009, that a client fire brigade in (b)(6) 2009 had observed an ignition of a liv oxygen. As (b)(6) crew put on the cylinder for service (first opening) sparks occurred around the bottom valve opening. The valve was closed, but sparks continued out of the bottom of the valve (assembly shroud and outlet connection). There were no injuries and there is no damage.
Patient Sequence No: 1, Text Type: D, B5


[10252472] The present case is one out of 9 incident reports in connection with 7 liv ignitions. Evaluation summary: root cause analysis summary: the cause of ignition with highest probability is ignition of hydrocarbons, lubricant and/or debris by adiabatic compression. Potential sources of hydrocarbons and lubricants; lubricant (fomblin) migration, (lubricant from gce manufacturing of valve). Leaching of contaminants (phthalates) from o-rings in valves (lubricant in solvent) and hydrocarbon oil. Leaching of contaminants (phthalates) from fill connector at filling plant at linde potential sources of debris. Debris from manufacturing process of valve at gce. Debris from filling plant at linde. Debris from wear, valves in use. The design (filter and location of components in the high pressure side of the valve) of the gce liv valve 5 micron filter is likely to cause more sensitivity to adiabatic compression heating than the gce oxygen design.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003900188-2009-00032
MDR Report Key2805492
Report Source01
Date Received2008-09-22
Date Mfgr Received2009-07-17
Date Added to Maude2012-10-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer Street575 MOUNTAIN AVENUE
Manufacturer CityMURRAY HILL NJ 07974
Manufacturer CountryUS
Manufacturer Postal07974
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGCE COMBILITE I VIPRS, 5 MICRO FILTER, LIV FOR OXYGEN
Generic NameNONE
Product CodeECX
Date Received2008-09-22
Model Number070137
Lot Number2980643
Device Expiration Date2017-10-30
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLINDE GAS NORTH AMERICA LLC
Manufacturer AddressMURRAY HILL NJ 07974 US 07974


Patients

Patient NumberTreatmentOutcomeDate
10 2008-09-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.