MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-10-26 for RELIANCE SYNERGY WASHER manufactured by Steris Canada Corporation.
        [10485054]
The steris service technician returned to the user facility to make repairs to the unit, ran a test cycle and confirmed the unit was operational. No further issues have been reported.
 Patient Sequence No: 1, Text Type: N, H10
        [20637519]
The user facility reported that water was leaking from their reliance synergy washer. No injuries or procedural delays/cancellations were reported. Some ceiling tiles on the lower level were damaged.
 Patient Sequence No: 1, Text Type: D, B5
        [20909807]
A steris service technician inspected the unit and found a crack on the sump, just under the steam coil outlet. The technician is scheduled to make repairs to the unit.
 Patient Sequence No: 1, Text Type: N, H10
| Report Number | 9680353-2012-00089 | 
| MDR Report Key | 2806716 | 
| Report Source | 06 | 
| Date Received | 2012-10-26 | 
| Date of Report | 2012-10-26 | 
| Date of Event | 2012-09-26 | 
| Date Mfgr Received | 2012-09-26 | 
| Date Added to Maude | 2013-02-05 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 0 | 
| Event Location | 0 | 
| Manufacturer Contact | MS LINDSEY MCGOWAN | 
| Manufacturer Street | 5960 HEISLEY ROAD | 
| Manufacturer City | MENTOR OH 44060 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 44060 | 
| Manufacturer Phone | 4403927519 | 
| Manufacturer G1 | STERIS CANADA CORPORATION | 
| Manufacturer Street | 490, ARMAND-PARIS | 
| Manufacturer City | QUEBEC, QUEBEC GIC8A3 | 
| Manufacturer Country | CA | 
| Manufacturer Postal Code | GIC 8A3 | 
| Single Use | 3 | 
| Previous Use Code | 3 | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | RELIANCE SYNERGY WASHER | 
| Generic Name | WASHER | 
| Product Code | LDS | 
| Date Received | 2012-10-26 | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | Y | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | Y | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | STERIS CANADA CORPORATION | 
| Manufacturer Address | 490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2012-10-26 |