WAVESCAN WAVEFRONT SYSTEM ACV 0070-1478

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-10-30 for WAVESCAN WAVEFRONT SYSTEM ACV 0070-1478 manufactured by Abbott Medical Optics.

Event Text Entries

[21754813] The customer reported that when they turned on their system, smoke was observed coming out the back of the computer. The system was immediately turned off and disconnected.
Patient Sequence No: 1, Text Type: D, B5


[21886497] An amo field service engineer inspected the system at the customer's location and did not observe any evidence of smoke or fire. The field service engineer found that the power supply board was not functioning and replaced the power supply board and also performed calibrations. The system met specifications following repair and was placed back into use. All pertinent information available to amo has been submitted.?
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006695864-2012-00173
MDR Report Key2808978
Report Source05,06
Date Received2012-10-30
Date of Report2012-10-04
Date of Event2012-10-04
Date Mfgr Received2012-10-04
Device Manufacturer Date2006-12-29
Date Added to Maude2012-10-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCAROL KAIL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478598
Manufacturer G1ABBOTT MEDICAL OPTICS INC.
Manufacturer StreetAMO MANUFACTURING 510 COTTONWOOD DRIVE
Manufacturer CityMILPITAS CA 95035
Manufacturer CountryUS
Manufacturer Postal Code95035
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWAVESCAN WAVEFRONT SYSTEM
Generic NameREFRACTIVE MEASUREMENT
Product CodeHKO
Date Received2012-10-30
Model NumberACV
Catalog Number0070-1478
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MEDICAL OPTICS
Manufacturer AddressSANTA ANA CA


Patients

Patient NumberTreatmentOutcomeDate
10 2012-10-30