SYSTEM 1000 S1000L3P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,04 report with the FDA on 2012-11-01 for SYSTEM 1000 S1000L3P manufactured by Baxter Healthcare - Largo.

Event Text Entries

[3021855] A customer contacted baxter (b)(4) regarding a burnt odor coming from a tina hemodialysis machine, during citric heat clean, while the home patient (hp) was not connected. There was no patient injury or medical intervention.
Patient Sequence No: 1, Text Type: D, B5


[10282932] (b)(4). This complaint for a report of "burnt odor" was confirmed and the root cause was determined to be due to a blockage of air inlet because of dust accumulation. No sample was returned for evaluation as the device was evaluated by the field service engineer (fse) at the customer location. During the evaluation, the reported condition of: "smelling burning smell" was confirmed but not duplicated by the fse. Visual inspection was performed and no issues were observed. As part of corrective action for reported problem; the fse cleared all of the dust in the fan input grills (internal filter was checked and was found to be clean) and performed self test. The device was evaluated and all necessary functional tests were performed and passed as per baxter? S standard operating policies and procedures.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2012-03218
MDR Report Key2814191
Report Source01,04
Date Received2012-11-01
Date of Report2012-10-04
Date of Event2012-10-04
Date Mfgr Received2012-10-04
Date Added to Maude2013-02-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactVIRGINIA MCKINNEY
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone8472702835
Manufacturer G1BAXTER HEALTHCARE - LARGO
Manufacturer Street7511 114TH AVE. NORTH
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal Code33773
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYSTEM 1000
Generic NameSYSTEM, HEMODIALYSIS, ACCESS RECIRCULATION MONITORING
Product CodeMQS
Date Received2012-11-01
Catalog NumberS1000L3P
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - LARGO
Manufacturer Address7511 114TH AVE. NORTH LARGO FL 33773 US 33773


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.