PRINTER, THERMAL, 110V 11-781

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,05 report with the FDA on 2012-10-24 for PRINTER, THERMAL, 110V 11-781 manufactured by Alere San Diego, Inc..

Event Text Entries

[15152964] Caller reported ldx printer is burning labels. There is a burning odor and the labels are brown. The printer is hot to the touch. No injuries were reported.
Patient Sequence No: 1, Text Type: D, B5


[15293697] Investigation was performed on returned printer and power supply. Conclusion: customer's complaint was not replicated. No burning, no charring and no discoloration was observed on any component. Printer and power supply that was tested did not feel hot to the touch during or after investigation, and no smoking was observed. Contamination was found on roller, causing it not to move properly. Roller and printer worked properly after some of the contamination on roller was removed. No burning marks observed between roller and heater plate. Some faint printing was observed on the left side of the labels, which may have been caused by contamination on roller. No burning marks or brown marks observed on the labels printed during investigation. Issue with the contamination observed on roller cannot be ruled out as the cause of the customer's observations. Contamination may cause the roller and labels that are being fed through to get stuck and cause the printer to print more info on the part of the label stuck between the roller and heater. No corrective action required at this time as no product deficiency was established.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2012-01557
MDR Report Key2824201
Report Source00,05
Date Received2012-10-24
Date of Report2012-10-24
Date of Event2012-10-10
Date Mfgr Received2012-10-10
Date Added to Maude2013-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCARMEN BERGELIN, MANAGER
Manufacturer Street9975 SUMMERS RIDGE RD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052256
Single Use3
Previous Use Code3
Removal Correction NumberNI
Event Type3
Type of Report3

Device Details

Brand NamePRINTER, THERMAL, 110V
Generic NameCHOLESTEROL METER
Product CodeCHH
Date Received2012-10-24
Returned To Mfg2012-10-15
Model Number11-781
Lot NumberNG
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2012-10-24