IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-11-09 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2936373] A customer reported to beckman coulter, inc. (bec) that the sample wash station on immage 800 immunochemistry system overflowed during daily set-up procedure. Approximately 1 ml of fluid overflowed and was contained within the instrument. No error codes were generated. The customer attempted to clear the wash drain/station with sodium peroxide, but the issue persisted. The customer was wearing a lab coat, gloves and eye protection at the time of the event. There was no report of injury or exposure, and medical attention was not sought. Msds was not reviewed, but the facility has an exposure control or risk management plan in place. No erroneous patient results were generated. Bec field service engineer (fse) advised the customer over the phone that the customer identified and cleared an occlusion in the wash station. The customer then cleaned the wash station with bleach.
Patient Sequence No: 1, Text Type: D, B5


[10309018] Obstruction in the sample wash station. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-01817
MDR Report Key2826039
Report Source05,06
Date Received2012-11-09
Date of Report2012-10-15
Date of Event2012-10-15
Date Mfgr Received2012-10-15
Device Manufacturer Date2008-03-14
Date Added to Maude2013-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2012-11-09
Model NumberIMMAGE 800
Catalog NumberA15445
Lot NumberNA
ID NumberSOFTWARE V. 1.6
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.