IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-11-12 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2938741] The customer reported an uncontained fluid leak underneath the immage 800 immunochemistry systems, which leaked onto the counter top. Per customer this is a new instrument and they have not run patient samples only quality controls on it. The customer was wearing personal protective equipment (ppe) consisting of a lab coat, eye protection, and gloves. The customer did not come in contact with the fluid and did not seek medical attention. No injury or exposure (sprayed or splashed) to mucous membranes or open wounds were reported.
Patient Sequence No: 1, Text Type: D, B5


[10300545] On (b)(4) 20128, a field service engineer (fse) was on site and found a loose connection between the wash buffer supply and the syringe selector valve. The fse secured the fitting internal probe wash assembly. The failure mode of leak was the loose fitting internal prove wash assembly. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-01821
MDR Report Key2828157
Report Source05,06
Date Received2012-11-12
Date of Report2012-10-17
Date of Event2012-10-17
Date Mfgr Received2012-10-17
Device Manufacturer Date2012-07-26
Date Added to Maude2013-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA, CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2012-11-12
Catalog NumberA15445
Lot NumberN/A
ID NumberSW 4.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD BREA, CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.