[15425932] Our newly installed nurse call sysem (responder 5) has been causing interference with our infant security system (safe place infant security system) causing multiple false alarms. The nurse call system is emitting noise interference in the 260-370 mgh level. The infant security system was installed many yrs ago and has been reliable, however when the new nurse call system was powered-up, alarms began immediately.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5027558
MDR Report Key2830257
Date Received2012-11-06
Date of Report2012-11-06
Date of Event2012-10-01
Date Added to Maude2012-11-14
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameRAULAND-BORG
Generic NameNURSE CALL SYSTEM
Product CodeILQ
Date Received2012-11-06
OperatorHEALTH PROFESSIONAL
Device Sequence No1
Device Event Key0
ManufacturerRAULAND-BORG