IMMULITE 2000 IGF-1 L2KGF2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-11-16 for IMMULITE 2000 IGF-1 L2KGF2 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[10496395] The initial mdr 2432235-2012-00387 was filed on (b)(4) 2012. (b)(4) 2012: additional information: the corrections and removal report (crr) was filed with the fda on (b)(4) 2012. The crr number is 2432235-11/28/2012-007-c.
Patient Sequence No: 1, Text Type: N, H10


[18313174] Falsely elevated insulin-like growth factor (igf-1) results were obtained on a patient sample on an immulite 2000 instrument. The patient's igf-1 results have been increasing over time. The results were reported to the physician(s), who questioned the results because they do not fit the clinical picture of the patient. There were no reports of patient intervention or adverse health consequences due to the discordant igf-1 results.
Patient Sequence No: 1, Text Type: D, B5


[18407800] The cause of the falsely elevated igf-1 results is unknown. An urgent field safety notice (ufsn), (b)(4)- "immulite / immulite 1000 / immulite 2000 / immulite 2000 xpi all immulite platforms for igf-i shift in patient medians and supply disruption" was sent to customers in (b)(6) 2012. Siemens is investigating this issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2012-00387
MDR Report Key2836438
Report Source05,06
Date Received2012-11-16
Date of Report2012-10-22
Date of Event2012-04-30
Date Mfgr Received2012-11-28
Date Added to Maude2013-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMINDY LOSAPIO
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242312
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer StreetGLYN RHONWY LLANBERIS, CAERNARFON
Manufacturer CityGWYNEDD, WALES LL554EL
Manufacturer CountryUK
Manufacturer Postal CodeLL55 4EL
Single Use3
Previous Use Code3
Removal Correction Number2432235-11/28/2012-007-C
Event Type3
Type of Report3

Device Details

Brand NameIMMULITE 2000 IGF-1
Generic NameIMMULITE 2000
Product CodeCFL
Date Received2012-11-16
Model NumberIMMULITE 2000 IGF-1
Catalog NumberL2KGF2
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer AddressGLYN RHONWY LLANBERIS, CAERNARFON GWYNEDD, WALES LL554EL UK LL55 4EL


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.