IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM N/A A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-11-19 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM N/A A15445 manufactured by Beckman Coulter Inc..

Event Text Entries

[21020647] The customer reported that the sample wash station of the immage 800 immunochemistry system was intermittently overflowing. The customer indicated that 1 ml of fluid had overflowed while in normal operating mode and no vacuum errors were generated. The customer was wearing personal protective equipment consisting of a lab coat, gloves and eye protection and no exposure or injury was reported. The customer stated that no erroneous results were generated and there was no change to patient treatment in connection with the event. Beckman coulter field service engineer (fse) was dispatched and found a loose fitting on the vacuum line going to the cup. The fse secured the fitting and repair was verified per established procedures. Failure mode was determined to be a loose fitting.
Patient Sequence No: 1, Text Type: D, B5


[21309978] Results: loose fitting. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-01862
MDR Report Key2838911
Report Source05,06
Date Received2012-11-19
Date of Report2012-10-31
Date of Event2012-10-31
Date Mfgr Received2012-10-31
Device Manufacturer Date2008-09-23
Date Added to Maude2013-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2012-11-19
Model NumberN/A
Catalog NumberA15445
ID NumberSW VERSION 1.9
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.