MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-11-20 for IMMULITE 2000 IGF-1 L2KGF2 manufactured by Siemens Healthcare Diagnostics.
[10463917]
The initial mdr 2432235-2012-00384 was filed on (b)(4), 2012. (b)(4) 2012: additional information: the corrections and removal report (crr) was filed with the fda on (b)(4), 2012. The crr number is 2432235-11/28/2012-007-c.
Patient Sequence No: 1, Text Type: N, H10
[18623152]
Falsely elevated insulin-like growth factor (igf-1) results were obtained on a patient sample on an immulite 2000 instrument. The results were reported to the physician(s), who questioned the results. A sample from the patient had been tested at a different laboratory on a different platform approximately three months before, and the result had been lower. There were no reports of patient intervention or adverse health consequences due to the discordant igf-1 results.
Patient Sequence No: 1, Text Type: D, B5
[18754867]
The cause of the falsely elevated igf-1 results is unknown. An urgent field safety notice (ufsn), ufsn 4005 - "immulite / immulite 1000 / immulite 2000 / immulite 2000 xpi all immulite platforms for igf-i shift in patient medians and supply disruption" was sent to customers in november 2012. Siemens is investigating this issue.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2432235-2012-00384 |
MDR Report Key | 2839778 |
Report Source | 05,06 |
Date Received | 2012-11-20 |
Date of Report | 2012-10-22 |
Date of Event | 2012-08-21 |
Date Mfgr Received | 2012-11-28 |
Date Added to Maude | 2013-02-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MINDY LOSAPIO |
Manufacturer Street | 511 BENEDICT AVENUE |
Manufacturer City | TARRYTOWN NY 10591 |
Manufacturer Country | US |
Manufacturer Postal | 10591 |
Manufacturer Phone | 9145242312 |
Manufacturer G1 | SIEMENS HEALTHCARE DIAGNOSTICS |
Manufacturer Street | GLYN RHONWY LLANBERIS, CAERNARFON |
Manufacturer City | GWYNEDD, WALES LL554EL |
Manufacturer Country | UK |
Manufacturer Postal Code | LL55 4EL |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | 2432235-11/28/2012-007-C |
Event Type | 3 |
Type of Report | 3 |
Brand Name | IMMULITE 2000 IGF-1 |
Generic Name | IMMULITE 2000 |
Product Code | CFL |
Date Received | 2012-11-20 |
Model Number | IMMULITE 2000 IGF-1 |
Catalog Number | L2KGF2 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS |
Manufacturer Address | GLYN RHONWY LLANBERIS, CAERNARFON GWYNEDD, WALES LL554EL UK LL55 4EL |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2012-11-20 |